Print Back
Western Innovator AGM2 Manifolds are designed and manufactured according to NFPA-99 2012 safety and performance requirements. Easy-to-read analog gauges indicate the line and individual bank pressures. A green light means the service bank is functioning and the reserve bank is ready for service. A red light alerts the user that one or both banks are depleted. A simple rotation of the control lever resets the unit.
AGM2 Series Carbon Dioxide Gas Manifold Systems
AGM2 Series Carbon Dioxide Gas Manifold Systems - 2
Unit of Measure
Items AGM2 Series Carbon Dioxide Gas Manifold Systems
AGM2HL-4-4
500 scfh with Heater Maximum Flow Rate, Carbon Dioxide Gas, and 4 Cylinders Manifold System with Standard 10 in. on Center and Wall mount
AGM2 Series Carbon Dioxide Gas Manifold Systems
AGM2HL-4-4-F
500 scfh with Heater Maximum Flow Rate, Carbon Dioxide Gas, and 4 Cylinders Manifold System with Standard 10 in. on Center and Floor Mounted
AGM2 Series Carbon Dioxide Gas Manifold Systems
AGM2HL-4-4-V
500 scfh with Heater Maximum Flow Rate, Carbon Dioxide Gas, and 4 Cylinders Manifold System with Vertical crossover 10 in. on Center and Wall mount
AGM2 Series Carbon Dioxide Gas Manifold Systems
AGM2HL-4-4-V-F
500 scfh with Heater Maximum Flow Rate, Carbon Dioxide Gas, and 4 Cylinders Manifold System with Vertical crossover 10 in. on Center and Floor Mounted
AGM2 Series Carbon Dioxide Gas Manifold Systems
AGM2HL-4-4-S
500 scfh with Heater Maximum Flow Rate, Carbon Dioxide Gas, and 4 Cylinders Manifold System with Staggered 5 in. on Center and Wall mount
Description N/A Western Innovator AGM2 Manifolds are designed and manufactured according to NFPA-99 2012 safety and performance requirements. Easy-to-read analog gauges indicate the line and individual bank pressures. A green light means the service bank is functioning and the reserve bank is ready for service. A red light alerts the user that one or both banks are depleted. A simple rotation of the control lever resets the unit.
Field N/A Medical
Maximum Inlet Pressure N/A 2000 psig
Maximum Flow Rate N/A With Heater: 500 scfh
Internal Adjustable Dual Line Regulators N/A 30 to 55 psig
Manifold Outlet Size N/A 1/2 in
Manifold Outlet Thread Type N/A Male National Pipe Thread (NPT)
Relief Valve Outlet Size N/A 1/2 in
Relief Valve Outlet Thread Type N/A Male National Pipe Thread (MNPT)
Overall Manifold Length N/A 5 ft 11 in. 1.80 m N/A 5 ft 11 in. 1.80 m N/A 4 ft 3 in. 1.30 m N/A 4 ft 3 in. 1.30 m N/A 5 ft 1 in. 1.55 m
Gas Service N/A Carbon Dioxide CGA-320
Number of Cylinders N/A 4
Header Configuration N/A Standard N/A Standard N/A Vertical Crossover N/A Vertical Crossover N/A Staggered
Mounting N/A Wall Mount N/A Floor Mounted N/A Wall Mount N/A Floor Mounted N/A Wall Mount
Additional Information N/A
  • NFPA Compliant, Automatic Analog System
  • CSA Approved Power Supply included
  • Oxygen Pigtails: NFPA 99 compliant, rigid copper, with check valves
  • Other Gas Pigtails: NFPA 99 compliant, 24" flexible stainless steel with check valves
  • 1/2" Brass, silver brazed headers
  • Digital readout can display psig, kPa or Bar
  • Connects to remote alarms systems (Up to 3 amps 30 VDC or 2 amps 250 VAC)
Warning N/A This product can expose you to lead, which is known to the State of California to cause cancer and birth defects or other reproductive harm.