Print Back
The Western Innovator HBCS automatic manifold is designed for High Purity gas applications where a constant uninterrupted supply of gas is critical while maintaining gas purity. The HBCS may be configured to meet almost any lab application requirements. The compact manifold headers allow multiple gas cylinders, up to 4 per bank, to supply gas to the manifold. In order to maintain system purity and provide isolation, the headers also include a packless diaphragm valve. All systems include an adjustable outlet regulator with a choice of outlet pressure ranges.
Unit of Measure

Specifications

Field

N/A Medical

Outlet Pressure

N/A 0 to 50 psig

Intermediate Pressure Gauge

N/A 1-3/4 in

Maximum Inlet Pressure

N/A 3000 psig

Gas Service

N/A Air CGA-346

Additional Information

Additional Information

N/A
  • High Purity Automatic Analog System
  • Chrome plated brass bar stock regulators
  • 316L Stainless steel diaphragms
  • Designed for gas purity levels up to grade 5.5 (99.9995%)
  • Switchover pressure
    • R to L-180 psig
    • L to R-160 psig
  • PTFE seats and seals
  • 10 micron sintered bronze filters

Features

Features

N/A
  • 180 degree lever indicates active cylinder side
  • 2" dual scale inlet pressure and outlet pressure gauges
  • External threaded relief valve
  • Stainless steel wall mounting panel

Prop 65

Warning

N/A This product can expose you to lead, which is known to the State of California to cause cancer and birth defects or other reproductive harm.

Note

Note

N/A No header kit is needed for a standard 1 x 1 configuration